Operation
Multi 3410
87
ba75863y18
01/2015
Settings
The settings are made in the menu for the measured parameter, turbid-
ity. To open the settings, display the required measured parameter in
the measured value display and press the
<MENU/ENTER>
key. After
completing the settings, switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor.
Default settings are printed in
bold
.
Setting menu
VisoTurb
®
900-P
4.10.4 Calibration
Why calibrate?
The calibration line of the sensor is determined and stored through cal-
ibrating.
When to calibrate?
When the calibration interval has expired
at regular intervals
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record
of the last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the last calibration
records (max. 10)
Calibration
/
Calibration data
storage
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration
data to a connected USB mem-
ory device/USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data
storage to the interface
Calibration
/
Calibration interval
1 ...
30
...
999 d
Calibration interval
for the tur-
bidity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Resolution
0.1
1
Resolution of the FNU/NTU dis-
play
Reset
-
Resets all sensor settings to the
delivery condition (see section
4.14.1).
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......