Operation
Multi 3410
65
ba75863y18
01/2015
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.8.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview
The following settings are possible for D.O. sensors:
Calibration record (display)
Calibration data storage (display)
Calibration interval
FDO Check
Check interval
Salinity correction
Salinity (salinity equivalent)
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the
<MENU/ENTER>
key. After
completing the settings, switch to the measured value display with
<M>
.
Measured
parameter
Time interval
Stability in the time interval
D.O. concen-
tration
20 seconds
Δ
: Better than 0.03 mg/l
D.O. saturation
20 seconds
Δ
: Better than 0.4 %
D.O. partial
pressure
20 seconds
Δ
: Better than 0.8 mbar
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible
setting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration record of the
last calibration
Calibration
/
Cali-
bration data
storage
-
Displays the last calibration records
(max. 10)
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......