Multi 3430
Operation
82
ba75865e19
04/2015
Setting menu of
LR 925/01
Menu item
Possible set-
ting
Description
Cell constant
0.090 ... ...
0.100
...
0.110 cm
-1
The cell constant can be dis-
played and set here
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature
compensation (see section
4.10.3).
This setting is only available
for the measured parame-
ters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear tem-
perature compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured parame-
ters,
ϰ
and
ρ
.
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.16.1)