Multi 3430
Operation
40
ba75865e19
04/2015
interface without the AutoRead info.
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.7.3
Settings for pH measurements
Overview
The following settings are possible for pH measurements:
Calibration record (display)
Calibration data storage (display)
Calibration interval
Buffers for calibration
Single-point calibration
Unit for slope
Resolution
Settings
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the
<MENU/ENTER>
key. After completing the settings, switch to the mea-
sured value display with
<M>
.
Default settings are printed in
bold
.
3
Release the frozen measured value again with
<AR>
or
<M>
.
The [AR] status display disappears. The display switches back
to the previous indication.
Measured
parameter
Time interval
Stability in the time interval
pH value
15 seconds
Δ
: Better than 0.01 pH
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible setting
Description
Calibration
/
Calibration record
-
Displays the calibration record of the last cal-
ibration
Calibration
/
Calibration data
storage
-
Displays the last calibration records (max.
10)
Summary of Contents for WTW Multi 3430
Page 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Page 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Page 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Page 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Page 133: ......