Na 800
Commissioning
ba77237e02
12/2021
15
Overview
Commissioning
1
Pull off the watering cap.
It contains storing solution (sodium chloride solution 0.1 mol/l).
2
Electrodes that were stored dry have to be watered in storing solu-
tion for 24 hours.
Refill any missing 3 M potassium chloride solution in the electrolyte
compartment of the reference system.
The filling level of the electrolyte solution should always be at least 5
cm above the level of the test sample.
3
Rinse the electrode with deionized water.
4
Wipe the shaft using a clean paper towel.
The electrode is ready to measure.
1
2
3
4
6
5
1 Protective cap
2 S7 plug-in connector
3 Refilling openings
4 Shaft
5 Junction
6 Glass membrane