NitraVis 70x IQ TS
Measurement / Operation
31
ba75952e03
05/2017
Preparing the
sensor check or
Zero adjustment
Prepare the sensor check or
Zero adjustment
as follows:
1.
Switch to the measured value display with <M>.
2.
Use <
> to select the NitraVis 70x IQ TS sensor.
3.
Call up calibration with <C>. The
Maintenance condition:
Linked outputs are frozen.
window appears.
By activating the maintenance condition, linked outputs remain in
their present state. The cleaning and the determination of measured
values are switched off. The measured value display shows a flash
-
ing, invalid measured value ("----").
4.
Confirm the maintenance condition with <OK>.
5.
Take the sensor out of the test sample.
6.
Remove the blind plug or unscrew the compressed air connection from
the sensor.
7.
Thoroughly clean the sensor enclosure, sensor sleeve and the com
-
pressed air channels (see section 5.1 M
).
To finish the cleaning procedure, thoroughly rinse the entire sensor with
ultrapure water.
8.
Mount the cleaned sensor sleeve:
Loosen both coupling rings on the sensor sleeve.
Put the sensor sleeve on the sensor and align it:
– The measuring gap is in the center of the sensor sleeve
– The filling opening of the sensor sleeve and the measuring gap
opening of the sensor are offset by 90°.
Tighten both coupling rings on the sensor sleeve by hand.