4.4.4
D.O. calibration
Why calibrate?
D.O. sensors age. This changes the slope of the D.O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.
When to calibrate?
After connecting another D.O. sensor
When the sensor symbol flashes (after the calibration interval has
expired).
Calibration
/
Calibration interval
1 ... 999 d
Calibration interval
for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.
Sal correction
On
Off
Manual salt content correc-
tion for concentration mea-
surements.
Salinity
0.0 ... 70.0
Salinity or salinity equivalent
for the salt content correc-
tion.
This function is only avail-
able for concentration mea-
surements if the manual salt
content correction is
switched on.
Stability control
On
Off
Switches on or off the auto-
matic stability control during
measurement (see section
4.3.3)
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are
displayed with the selected
unit.
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.6.1).
Menu item
Possible
setting
Explanation
Summary of Contents for WTW Oxi 3210
Page 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Page 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Page 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Page 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Page 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Page 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Page 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...
Page 51: ......