Oxi 3210
Technical data
45
ba75795e07
10/2021
7.2
Measuring ranges, resolution, accuracy
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the D.O. sensors has also to be taken into account.
Measuring ranges,
resolution
(depending on the sen-
sor)
Note: The values quoted in brackets apply especially for the
DurOx 325 sensor.
Variable
Measuring range
Resolution
Concentration [mg/l]
0 ... 20.00 (0 ... 20.0)
0 ... 90.0 (0 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 200,0 (0 ... 200)
0 ... 600
0.1 (1)
1
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 200)
0 ... 1250
0.1 (1)
1
T [°C]
0 ... 50,0
0.1
Accuracy (± 1 digit)
Variable
Accuracy
Concentration [mg/l]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
Saturation [%]
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
T [°C]
/ temperature sensor
NTC 30
± 0.1
PT 1000
± 0.1
Correction functions
Temperature
compensation
Accuracy better than 2 % at 0 ... + 40 °C
Salinity correction
0 ... 70.0 SAL
Air pressure correc-
tion
Automatic through integrated
pressure sensor in the range of 500 ... 1100
mbar
Summary of Contents for WTW Oxi 3210
Page 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Page 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Page 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Page 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Page 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Page 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Page 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...