4.3
Sensor-independent settings
The
Storage & config
menu comprises the following settings:
System
Data storage
4.3.1
System
Overview
The following sensor-independent meter characteristics can be
adjusted in the
Storage & config
/
System
menu:
Menu language
Beep on keystroke
Illumination
Display contrast
Interval of the automatic switch-off function
Data interface
Clock and date function
Reset of all sensor-independent system settings to the default con-
dition
Settings
To open the
Storage & config
menu, press the
<F1
_
>
/[Menu] key in
the measured value display. After completing the settings, switch to the
measured value display with
<M>
.
Menu item
Setting
Explanation
System
/
General
/
Language
Deutsch
English
(more)
Selects the menu lan-
guage
System
/
General
/
Beep
On
Off
Switches on/off the beep
on keystroke
System
/
General
/
Illumination
Auto
On
Off
Switches the display illumi-
nation on/off
System
/
General
/
Contrast
0 ... 100 %
Changes the display con-
trast
System
/
General
/
Switchoff time
10 min ... 24 h
Adjusts the switch-off time
System
/
Interface
/
Baud rate
1200, 2400,
4800, 9600,
19200
Baud rate of the data inter-
face
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......