Operation
Oxi 3310
28
ba75804e08
10/2017
4.4.2
Measuring
You can carry out D.O. measurements as follows:
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction
When measuring the concentration of solutions with a salt content of
more than 1 g/l, a salinity correction is required. For this, you have to
measure and input the salinity of the measured medium first. When the
salinity correction is switched on, the
[SAL]
indicator is displayed in the
measuring window.
Note
You can switch the salinity correction on or off and enter the salinity in
the menu for calibration and measurement settings (see section 4.4.3).
Stability control
(AutoRead)
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic
Stability control
is switched off.
Irrespective of the setting for automatic
Stability control
in the
Measurement
menu, you can start the
Stability control
function
manually at any time.
1
Perform the preparatory activities according to section 4.4.1.
2
Immerse the D.O. sensor in the test sample.
O
2
7.92
mg/l
25.1
°C
[SAL]
15.03.2014
08:00
Menu
USB output
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......