background image

Oxi 3310

Operation

37

ba75804e08

10/2017

4.5

Data memory

You can transmit measured values (datasets) to the data memory:

Manual storage (see section 4.5.1)

Automatic storing at intervals (see section 4.5.2)

Each data storing process transmits the current dataset to the interface 
at the same time.

Measurement dataset

A complete dataset consists of:

ID number

Date/time

Measured value of the connected sensor

Measured temperature value of the connected sensor

AutoRead info: 

AR

 appears with the measured value if the 

AutoRead criterion was met while storing (stable measured value). 
Otherwise, the 

AR

 display is missing.

Calibration evaluation: +++, ++, +, -, or no evaluation

Memory locations

The Oxi 3310 meter has two measurement data memories. The mea-
sured values recorded either manually or automatic are stored sepa-
rately in individual measurement data memories.

Data memory

Maximum number of datasets

Manual data storage

500

Automatic data storage

5000

Summary of Contents for WTW Oxi 3310

Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...

Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...modes 18 4 2 2 Navigation 19 4 2 3 Navigation example 1 Setting the language 21 4 2 4 Example 2 on navigation Setting the date and time 23 4 3 Sensor independent settings 25 4 3 1 System 25 4 3 2 Data...

Page 4: ...4 8 Reset 46 4 8 1 Resetting the measurement settings 46 4 8 2 Resetting the system settings 47 5 Maintenance cleaning disposal 49 5 1 Maintenance 49 5 1 1 Replacing the batteries 49 5 2 Cleaning 50...

Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...

Page 6: ...and reliably The Oxi 3310 provides the maximum degree of operating comfort reli ability and measuring certainty for all applications The proven OxiCal calibration procedure supports you in your work...

Page 7: ...menu for system settings On Off Switches the meter on or off M Selects the measured parameter CAL CAL_ Calls up the calibration procedure Displays the calibration data STO STO_ Saves a measured value...

Page 8: ...ensor symbol calibration evaluation calibration interval 5 Measured temperature with unit 6 Status line 7 Softkeys and date time 3 2 4 5 6 7 1 HOLD AR 25 0 7 92 C O2 mg l 15 03 2014 08 00 Menu USB out...

Page 9: ...connect D O sensors to the meter that cannot return any voltages or currents that are not allowed SELV and current circuit with current limiting Almost all customary D O sensors fulfill these conditio...

Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...

Page 11: ...necessary safety precautions to take when handling chemicals Safety instructions Safety instructions in this operating manual are indicated by the warn ing symbol triangle in the left column The signa...

Page 12: ...al safety The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual ar...

Page 13: ...ou are in any doubt please contact the supplier of the meter Obligations of the purchaser The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangero...

Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...

Page 15: ...ith USB drivers detailed operating manual Software MultiLab Importer 3 2 Initial commissioning Perform the following activities Insert the supplied batteries Switch on the meter Set the date and time...

Page 16: ...pletion The energy saving feature switches off the meter if no key is pressed during the adjusted interval How to set the switch off interval see sec tion 4 3 1 3 2 3 Setting the date and time 1 Place...

Page 17: ...4 3 1 The automatic switchoff function switches off the meter if no key is pressed for an adjustable period The automatic switchoff function is not active if the meter is supplied with power via a co...

Page 18: ...1 and section 4 2 2 4 2 1 Operating modes The instrument has the following operating modes Measuring The measurement data of the connected sensor are shown in the measured value display Calibration Th...

Page 19: ...ings and dialogs in procedures contain further sub elements The selection is made with the keys The current selection is displayed with a frame Submenus The name of the submenu is displayed at the upp...

Page 20: ...cannot be selected Example Note The principles of navigation are explained in the two following sections by reference of examples Setting the language section 4 2 3 Setting the date and time see secti...

Page 21: ...e instrument is in the measuring mode 2 Using F1_ Menu open the Storage config menu The instrument is in the setting mode 3 Select the System submenu with The current selection is displayed with a fra...

Page 22: ...e setting mode for the Language with ENTER 8 Select the required language with General Interface Clock Service information Reset System 15 03 2014 08 00 Back Language Deutsch Beep Off Illumination On...

Page 23: ...alues Therefore check the time at regular intervals Note After a fall of the supply voltage empty batteries the date and time are reset Setting the date time and date format The date format can be swi...

Page 24: ...nge and confirm the setting with and ENTER The time is set 7 If necessary set the Date and Date format The setting is made similarly to that of the time 8 To make further settings switch to the next h...

Page 25: ...ependent system settings to the default con dition Settings To open the Storage config menu press the F1_ Menu key in the measured value display After completing the settings switch to the measured va...

Page 26: ...on 4 4 3 The measured parameter flashes on the display as soon as the measured value is outside the stability range when you switch over between the measured parameters with M when the automatic Stabi...

Page 27: ...n When a grounded PC is connected measurements cannot be per formed in grounded media as incorrect values would result The USB interface is not galvanically isolated Temperature measurement The D O se...

Page 28: ...ty correction on or off and enter the salinity in the menu for calibration and measurement settings see section 4 4 3 Stability control AutoRead The stability control function AutoRead continually che...

Page 29: ...arameter flashes The HOLD AR status indicator appears as soon as a stable measured value is recognized Note You can prematurely terminate the Stability control function manually with ENTER at any time...

Page 30: ...tic Stability control Settings The settings are available in the menu for measurement and calibration settings To open the settings display the required parameter in the measured value display and pre...

Page 31: ...ement e g Winkler titration For details see section 4 4 4 Sal correction On Off Manual salt content correc tion for concentration mea surements Salinity 0 0 70 0 Salinity or salinity equivalent for th...

Page 32: ...g to DIN EN 25813 or ISO 5813 At the same time the rel ative slope is adapted to the comparison measurement by a correc tion factor When the correction multiplier is active the Factor indicator appear...

Page 33: ...to calibrate the instrument Note The sponge in the air calibration vessel must be moist not wet Leave the sensor in the air calibration vessel for a time long enough to adjust Display Calibration reco...

Page 34: ...4 Start the measurement with ENTER The measured value is checked for stability stability control The AR status indicator is displayed The measured parame ter flashes 5 Wait for the end of the AutoRead...

Page 35: ...is checked for stability stability control The AR status indicator is displayed The measured parame ter flashes 5 Wait for the end of the AutoRead measurement or accept the calibration value with ENT...

Page 36: ...ata stor age Display Displays the calibration record Further options Scroll through the cali bration records with Output the displayed calibration record to the interface with F2 USB output Quit the d...

Page 37: ...sured value of the connected sensor Measured temperature value of the connected sensor AutoRead info AR appears with the measured value if the AutoRead criterion was met while storing stable measured...

Page 38: ...cess and switch to the measured value dis play confirm No Then you can e g store the data from the memory to a PC see section 4 5 3 and subsequently erase the memory see section 4 5 4 1 Press the STO...

Page 39: ...the STO_ key The menu for automatic data storing appears Specified entire storing duration Max available storing dura tion Graphical display of the memory usage ID number 1 Interval 30 s Duration 180...

Page 40: ...ode If the automatic storing function is active the meter provides an energy saving mode Eco mode to avoid unnecessary energy consumption The energy saving mode switches off functions of the meter tha...

Page 41: ...y can be shown on the display and output to the interface Editing the data memory The memory is edited in the menu Storage config Data storage To open the Storage config menu press the F1_ Menu key in...

Page 42: ...put the displayed dataset to the interface with F2 USB output Quit the display with F1 Back Data storage Manual data storage Erase Erases the entire manual measurement data memory Note All calibration...

Page 43: ...play and move to the next higher menu level with F1 Back 4 5 4 Erasing the measurement data memory How to erase the measurement data memory is described in section 4 5 3 DISPLAYING AND EDITING THE MEA...

Page 44: ...on records Data Control Operation description Currentmeasured values of all con nected D O sen sors Manual With F2 USB output Simultaneously with every manual storing process see section 4 5 1 Automat...

Page 45: ...and CD ROM drive Microsoft Windows for details see enclosed installation CD Driver directory 4 7 MultiLab Importer With the aid of the MultiLab Importer software you can record and eval uate measureme...

Page 46: ...a reset The following settings for D O measurements are reset to the default settings with the Reset function The sensor settings are reset under the Reset menu item in the menu for calibration and m...

Page 47: ...t to the delivery status The system settings are reset in the menu Storage config System Reset To open the Storage config menu press the F1_ Menu key in the measured value display Setting Default sett...

Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...

Page 49: ...maintenance 5 1 1 Replacing the batteries Caution Make sure that the poles of the batteries are positioned correctly The signs on the batteries must correspond to the signs in the battery compartment...

Page 50: ...etone or similar detergents that contain solvents Re move any splashes immediately 5 3 Packing This meter is sent out in a protective transport packing We recommend Keep the packing material The origi...

Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...

Page 52: ...sor contaminated Clean D O sensor and replace it if necessary Cause Remedy Cleaning interval expired Recalibrate the measuring system Cause Remedy Batteries almost empty Replace the batteries see sect...

Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...

Page 54: ...alkali manganese batteries type AA Rechargeable batter ies 4 x 1 2 V NiMH rechargeable batteries type AA no charging function Operational life Up to 800 h without 100 h with illumination USB interface...

Page 55: ...uration 0 200 0 0 200 0 600 0 1 1 1 D O partial pressure mbar 0 200 0 0 200 0 1250 0 1 1 1 T C 0 50 0 0 1 Accuracy 1 digit Variable Accuracy Concentration mg l 0 5 of measured value at ambient tempera...

Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...

Page 57: ...g system e g the displayed value to the correct value or a value that is regarded as correct Often this expression is also used when the measuring system is adjusted at the same time see adjusting D O...

Page 58: ...conductivity Salt content General designation for the quantity of salt dissolved in water Slope relative Designation used by WTW in the D O measuring technique It ex presses the relation of the slope...

Page 59: ...re update 61 I Initial commissioning 15 16 Initialize 46 K Keys 7 M Measured value display 19 Measurement data memory Edit 41 Erase 41 Memory locations 37 Measurement dataset 37 Measuring 28 Menu for...

Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...

Page 61: ...version see page 52 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter or meter type the new data...

Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...

Page 63: ......

Page 64: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: