Automatic storing at intervals
The storing interval (
Interval
) determines the time interval between
automatic data storing processes. Each data storing process transmits
the current dataset to the interface at the same time.
Configuring the
automatic memory
function
Settings
You can configure the automatic data storing function with the following
settings:
1
Press the
<STO
_
>
key.
The menu for automatic data storing appears.
Specified entire storing
duration
Max. available storing dura-
tion
Graphical display of the
memory usage
ID number
1
Interval
30 s
Duration
180 min
Continue
0d03h00min
0
1d17h33min
Automatic data storage
15.03.2014
08:00
Back
Menu item
Possible
setting
Explanation
ID number
1 ... 10000
ID number for the dataset
series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min,
30 min, 60 min
Storing interval.
The lower limit of the stor-
ing interval can be
restricted by the number of
free memory locations.
The upper limit is restricted
by the storing duration.
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......