What to do if...
Oxi 3310
52
ba75804e08
10/2017
6
What to do if...
Error message,
OFL
Error message,
Error
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause
Remedy
– Measured value outside the
measuring range
– Use a suitable D.O. sensor
Cause
Remedy
– D.O. sensor contaminated
– Clean D.O. sensor and
replace it if necessary
Cause
Remedy
– Cleaning interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultane-
ously
Cause
Remedy
– E.g., a question by the service
department
– Switch on the meter.
Open the menu,
<F1
_
>
/
[Menu] /
Storage & config
/
SystemService information
.
The instrument data are dis-
played.
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......