pH value
Multi 3630 IDS
32
ba77170e03
10/2017
2
Start the calibration with
<CAL>
.
The calibration display for the first buffer appears (voltage display).
3
Thoroughly rinse the IDS sensor with deionized water.
4
Immerse the IDS-pH sensor in buffer solution 1.
5
For measurements without temperature sensor
(e.g. when using an IDS adapter
)
:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
6
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
7
Wait for the end of the measurement with stability control or accept the
calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
8
If necessary, finish the calibration procedure as a single-point calibra-
tion with
<M>
.
The calibration record is displayed.
30.09.2015 08:00
30.09.2015 08:00