Turbidity measurement (VisoTurb
®
900-P)
Multi 3630 IDS
68
ba77170e03
10/2017
Continuing with
three-point
calibration
Repeat the steps 11 to 15 with the third standard solution. The new calibration
values are displayed after the last calibration step was completed.
9.2.5 Calibration data
Displays the
calibration data
Calibration
The calibration protocol of the last calibration is available in the
menu
<ENTER>
/
Calibration record
. To open it in the measured value display,
press the
<CAL_>
key.
The calibration records of the last 10 calibrations are available in the menu
Cal-
ibration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in
the measured value display, press the
<ENTER>
key.
14
Use
<
> <
>
and
<
><
>
to set the concentration of the standard
solution for each digit and confirm with
<ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
15
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
16
If necessary, terminate the calibration as a two-point calibration with
<M>
.
The new calibration values are displayed.
or
Continue with three-point calibration.
15.09.2016 08:00
1010.0 FNU
Standard
3
Cal