Multi 3630 IDS
Depth of immersion (multi parameter probe MPP 9x0 IDS)
ba77170e03
10/2017
73
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
10.2.2 Measuring the temperature
The multi parameter probe MPP 9x0 IDS measures the temperature with a tem-
perature sensor integrated in the probe.
This temperature is displayed if no IDS sensor of the XXX-P series is con-
nected to the multi parameter probe MPP 9x0 IDS.
As soon as an IDS sensor of the XXX-P series is connected to the multi param-
eter probe MPP 9x0 IDS, the temperature value of the IDS sensor of the XXX-
P series is displayed for the main parameter.
10.3 Calibration
The multi parameter probe MPP 9x0 IDS is calibration free.
The IDS sensors of the XXX-P series are calibrated while being directly con-
2
Using
<ENTER>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<ENTER>
at any time. If the
Stability control
function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3
Using
<ENTER>
, start a further measurement with
Stability control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured param-
eter
Time interval
Stability in the time inter-
val
Depth of immersion
DPT
10 seconds
∆
: better than 0.20 m