Settings
Multi 3630 IDS
82
ba77170e03
10/2017
11.4 Cond measurement settings
11.4.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<ENTER>
key. After completing the settings,
switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
TetraCon 925
,
LR 925/01
).
Default settings are printed in
bold
.
Salinity
0.0
... 70.0
Salinity or salinity equivalent for
the salt content correction.
This menu item is only available if
the automatic salt content correc-
tion is switched off and the manual
salt content correction is switched
on.
Response time t90
30
... 300
Response time of the signal filter
(in seconds).
A signal filter in the sensor reduces
the limits of variation of the mea-
sured value. The signal filter is
characterized by the response
time t90. This is the time after
which 90 % of a signal change is
displayed.
This menu item is only available if
sensor and meter support this
function.
You can update the firmware of
sensor and meter (see chapter
17).
Reset
-
Resets all sensor settings to the
delivery condition (see section
11.7.1).
Menu item
Possible
setting
Explanation