Multi 3630 IDS
Settings
ba77170e03
10/2017
89
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
Conductivity
The following settings for conductivity measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
Turbidity
The following settings for the turbidity measurement are reset to the default
condition with the
Reset
function:
Setting
Default settings
Calibration interval
180 d
Check interval
60 d
Measured parameter
D.O. concentration
Relative slope (S
Rel
)
1.00
Salinity (value)
0.0
Salinity (function)
Off
Setting
Default settings
Calibration interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of
the linear temperature compensa-
tion
2.000 %/K
TDS factor
1.00