pHotoFlex
®
Turb
Operation
39
ba75976e07
09/2021
4 Open the
Standard adjustment
menu with
<
V
> <
W
>
and
<START/
ENTER>
.
If data of a standard adjustment are already available, the data of
the last standard adjustment are displayed. Here you can also
erase the data of an active standard adjustment.
5 Using
<
V
> <
W
>
and
<START/ENTER>
, select and start the
Mea-
sure standard
function.
The menu-guided standard measurement begins.
Follow the instructions on the display.
6 Enter the nominal value of the standard.
Note:
Enter the decimal separator with
<START/ENTER>
.
7 Confirm the entered nominal value with
<START/ENTER>
.
8 Insert a cell with standard (see section 4.2).
9 Start the measurement of the standard with
<START/ENTER>
.
After measuring, the result of the standard adjustment is displayed
and stored.
As the result, the measured value and the adjustment (in %) or
Error
is displayed.
Zero adjust.
Measure blank value
Delete blank value
Standard adjustment
Photometry \ Adjustment
mg/l Cu
_
Enter nominal value