4.5.7 Zero adjustment
The zero adjustment, i. e. measuring and storing the absorbance of a cell
filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambient
temperature has changed.
Only perform the zero adjustment against distilled water in an optically per-
fect cell. The zero adjustment must be performed separately for each cell
type.
3 Select the cell diameter with
<
V
> <
W
>
and confirm with
<START/
ENTER>
.
4 Select the wavelength with
<
V
> <
W
>
and confirm with
<START/EN-
TER>
.
5 Clean the cell (see section 5.2.2).
6 Insert the cell (see section 4.2).
7 Start the measurement with
<START/ENTER>
.
The measurement result is displayed when the measurement is fin-
ished.
Photometry
\
Absorbance
i
Insert sample
i
Start measurement
with <START>
16 mm
610 nm
01.02.05 15:12
Photometry
\
Absorbance
0.532
16 mm
610 nm
01.02.05 15:12