What to do if...
pHotoFlex
®
Turb
82
ba75976e07
09/2021
6
What to do if...
6.1
General errors
Display,
LoBat
Meter does not
react to keystroke
RS232 interface
does not react
Error message,
Error
0, 8, 16, 16384
Cause
Remedy
– Batteries or battery pack
almost empty
– Insert new batteries
– Charge the battery pack(see
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
key simultaneously.
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
keys simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
Send in the meter for repair, stating
the error number