pHotoFlex
®
Turb
Operation
29
ba75976e07
09/2021
4.4.1
Measured value memory
In the
Measured value memory
menu, you find functions to display and edit
the stored measurement datasets:
Show measurement datasets on the display (
Display
)
Output measurement datasets to the RS232 interface (
RS232 download
)
Define filter rules for the stored measurement datasets (
Data filter
)
Erase all stored measurement datasets (
Delete
)
Information on the number of occupied memory locations
The settings are in the menu,
Configuration
/
System
/
Measured value mem-
ory
. Move to the
Configuration
menu with the
<MENU>
key.
Settings/functions
Continue ...
/
Switchoff time
10, 20, 30, 40,
50 min,
1, 2, 3, 4, 5, 10,
15, 20, 24 h
The automatic switchoff func-
tion switches the meter off if no
entry is made for a specified pe-
riod of time (
Switchoff time
).
This conserves the batteries or
the battery pack.
Continue ...
/
Beep
On
Off
Switches on/off the beep on
keystroke
Menu item
Setting
Explanation
Menu item
Setting/
function
Explanation
Display
-
Displays in pages all measurement data-
sets that correspond to the filter settings.
Further options:
Scroll through the datasets with
<
V
> <
W
>
.
Output the displayed dataset to the inter-
face with
<PRT>
.
Quit the display with
<ESC>
.
RS232 down-
load
-
Downloads to the interface all measure-
ment datasets that correspond to the filter
settings. The download is ordered accord-
ing to the date and time.
The process can take several minutes. To
terminate the process prematurely, press
<ESC>
.