Operation
pHotoFlex
®
Turb
44
ba75976e07
09/2021
program data for each program. The active
Analysis timer
function automat-
ically reminds you to observe these time intervals by means of the menu
guidance.
If you want to manually enter time intervals, use the
Timer
function (see
The
Analysis timer
with the required time interval is automatically displayed
at the due point.
Start the
Analysis timer
with the
<START/ENTER>
key.
Shortening the time interval is not possible.
An acoustic signal sounds when the adjusted time interval has expired.
The
Analysis timer
function is switched on or off in the
Configuration
/
Pho-
tometry
/
Analysis timer
menu.
This setting generally applies to all measurements with methods according to
analysis specification.
4.5.10
Timer
When measuring according to analysis specifications, waiting periods often
have to be kept between individual steps of the method.
With the
Timer
function you manually set a time interval.
If you want to be automatically reminded of the given time intervals, use the
Analysis timer
function (see section 4.5.9).
The timer is displayed in the measured value display. It always displays the
remaining time of the adjusted time interval.
When the adjusted time interval has expired, the timer indicates 00:00:00,
and an acoustic signal sounds.
The
Timer
function is started in the
Configuration
/
Timer
menu by entering a
time interval.
4.5.11 Measuring diluted samples
If the concentration of a test sample exceeds the measuring range of a meth-
od, you can dilute the sample by a factor 1 ... 99 so that the concentration of
the diluted test sample is within the measuring range of the method (see pho-
tometry analysis manual). Thus a valid measurement is possible.
After entering the factor for the dilution the meter converts the concentration
to that of the undiluted sample.
The display then indicates the measured value of the undiluted sample.
Entering the fac-
tor of the dilution
1 Select the program for which a dilution factor is to be entered.