Operation
pHotoFlex
®
Turb
48
ba75976e07
09/2021
AutoRead (drift
control)
The AutoRead function (drift control) continually checks the stability of the
measurement signal. The stability has a considerable impact on the repro-
ducibility of measured values. The display of the measured parameter flash-
es until a stable measured value is available.
Criteria
With identical measurement conditions, the following applies:
4.6.4 Settings for pH and ORP measurements
Overview
For pH and ORP measurements, the following settings are possible in the
menu
Configuration
/
pH & ORP
:
Meas. param.
Calibr. record
(display, print)
Select the calibration type
Enter the
Calibr. interval
Select
Unit for slope
Select
Temperature unit
Reset
Settings/functions
The settings are in the menu,
Configuration
/
pH & ORP
.
Move to the
Configuration
menu with the
<MENU>
key.
3 Press the
<M>
key (short pressure) repeatedly until the measured
parameter,
ORP
is selected.
Measured parame-
ter
Reproducibility
Response time
ORP
better than 1 mV
> 30 seconds
Menu item
Possible
setting
Explanation
Meas. param.
pH & ORP
mV
Calibration
/
Calibr. record
-
Display the calibration record of
the last calibration.
Calibration
/
Calibr. type
TEC
NIST/DIN
Buffer sets to be used for pH cal-
ibration.
For details, see section 4.6.5.