background image

Storage

SensoLyt

®

 900(-P)

6

ba75992s02

02/2017

measurement to the next by taking the following measures:

– Shortly rinse the calibration and sample beakers with the solution the 

beakers are to be filled with next.

– Between measurements, rinse the electrode with the solution that fol-

lows. Alternatively, you can also rinse the electrode with deionized 
water and then carefully dab it dry.

To measure in aqueous solutions, it is recommended to immerse the elec-
trode in a vertical or slightly tilted position.

Observe the correct depth of immersion and make sure the contact 
between the junction and test sample is thorough. The junction (hole junc-
tion) is in the area of the bottom end of the shaft.

For measurements in aqueous solutions, provide approximately the same 
stirring conditions for measuring as for calibrating.

NOTE

Prevent contact of the pH membrane to the beaker bottom to avoid 
scratches on the pH membrane.

Subsequent

calibrations

The frequency of subsequent calibrations depends on the application. The 
meters provide an option where you can enter a calibration interval. After the 
calibration interval has expired, the meter will automatically remind you of the 
due calibration.

3

Storage

During short

measuring breaks

Immerse the electrode in reference electrolyte (KCl 3 mol/L, Ag

+

 free).  Prior 

to the next measurement, shortly rinse the electrode with the test sample or 
deionized water.

Overnight or

longer

Put the clean electrode in the watering cap that is filled with reference elec-
trolyte (KCl 3 mol/L, Ag

+

 free).

NOTE

Do not store the electrode dry or in deionized water. 
The electrode could be permanently damaged by this. 
If the liquid in the watering cap has dried up, condition the electrode in 
reference electrolyte (KCl 3 mol/L, Ag+  free) for at least 24 hours.

Junction

Summary of Contents for WTW SensoLyt 900

Page 1: ...OPERATING MANUAL ba75992e02 02 2017 SensoLyt 900 P IDS pH ELECTRODE WITH POLYMER ELECTROLYTE OPERATING MANUAL...

Page 2: ...ba75992e02 02 2017 SensoLyt 900 P 2 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...toring the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated sen...

Page 4: ...ction Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure th...

Page 5: ...ring the sensor for measurement 2 3 Calibration and measurement General rules Avoid the carryover of any solution sample or buffer solution from one Please keep the watering cap It is required for the...

Page 6: ...uring as for calibrating NOTE Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane Subsequent calibrations The frequency of subsequent calibrations depends on...

Page 7: ...e contamination by rinsing with deionized water Other types of contamination have to be removed as follows while the contact time with the detergents should be kept as short as possible NOTE Hydrofluo...

Page 8: ...nge 0 60 C Membrane resistance at 25 C 500 MOhm Typical application Field Accuracy of the IDS measuring technique Measured parameter Accuracy 1 digit pH 0 004 U mV 0 2 T C 0 1 Pressure range at temper...

Page 9: ...nnection cable 3 m AS IDS 3 903 851 IDS connection cable 6 m AS IDS 6 903 852 IDS connection cable 10 m AS IDS 10 903 853 IDS connection cable 15 m AS IDS 15 903 854 IDS connection cable 20 m AS IDS 2...

Page 10: ...Wear parts and accessories SensoLyt 900 P 10 ba75992s02 02 2017...

Page 11: ......

Page 12: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: