Commissioning
Scope of delivery
Electrode SenTix
®
950 / 980 / Micro 900(-P)
Operating manual
Commissioning
The electrode is filled with reference electrolyte solution in the factory. Pre-
pare the electrode for measuring as follows:
Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop-
per or a slider.
The refilling opening must always be open during calibration and
measurement!
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
SenTix
®
950(-P): Remove any gas bubbles behind the pH membrane by
shaking.
Connect the electrode to the meter.
Please keep the watering cap. It is required for the electrode to
be stored. Always keep the watering cap clean.
SenTix
®
950 / 980 /
Micro 900
– via the sensor cable to a free IDS connec-
tor on the meter
SenTix
®
950 / 980 /
Micro 900-P
– via a connecting cable (accessory) to a
free IDS connector on the meter
or
– wireless via an IDS WLM-S adapter
(accessory) to a WLM-capable meter