Commissioning
Scope of delivery
Electrode SenTix
®
HW-T 900(-P
)
Operating manual
Commissioning
The electrode is filled with reference electrolyte solution in the factory. Pre-
pare the electrode for measuring as follows:
Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop-
per or a slider.
The refilling opening must always be open during calibration and
measurement!
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Connect the electrode to the meter
Calibrate the electrode according to the operating manual of the meter.
When doing so, observe the general rules on the following page.
Ground junction of
SenTix
®
HW-
T 900
(
-P
)
With the ground junction of the SenTix
®
HW-T 900 (-P
)
electrodes, the elec-
trode outflow can be adjusted with a thread:
Hold the electrode with the pH glass membrane pointing downward. Using
the provided special wrench, open the ground junction until some electro-
lyte solution can be seen to come out of the junction and all ground sur-
faces are completely moist.
Please keep the watering cap. It is required for the electrode to
be stored. Always keep the watering cap clean.