lows. Alternatively, you can also rinse the electrode with deionized 
water and then carefully dab it dry.

Immerse the electrode in the solution in a vertical or slightly tilted position.

Make sure the immersion depth is correct. The ground junction must be 
completely submersed in the solution. At the same time, the level of the 
reference electrolyte must be at least 2 cm above the level of the solution.

Provide approximately the same stirring conditions for measuring as for 
calibrating.

Subsequent

calibrations

The frequency of subsequent calibrations depends on the application. Many 
meters provide an option where you can enter a calibration interval. After the 
calibration interval has expired, the meter will automatically remind you of the 
due calibration.

3

Storage

During short

measuring breaks

With the refilling opening open, immerse the electrode in reference electro-
lyte (KCl 3 mol/L, Ag

+

 free). Observe the minimum depth of immersion. Prior 

to the next measurement, shortly rinse the electrode with the test sample or 
deionized water.

Overnight or

longer

Put the clean electrode in the watering cap that is filled with reference elec-
trolyte (KCl 3 mol/L, Ag

+

 free) and close the refilling opening. 

NOTE

pH electrodes must not be stored dry or in deionized water. The elec-
trode could be permanently damaged by this. If the liquid in the 
watering cap has dried up, condition the electrode in reference electro-
lyte (KCl 3 mol/L, Ag+ + free) for at least 24 hours.

4

Aging

pH electrodes are consumables. Every pH electrode undergoes a natural 
aging process. With aging, the responding behavior becomes slower and 
slower and the electrode slope and asymmetry change. Moreover, extreme 
operating conditions can considerably shorten the lifetime of the electrode.

Prevent contact of the pH membrane to the beaker bottom to 
avoid scratches on the pH membrane.

During longer storing periods, salt sediments may develop on the 
watering cap. They do not affect the measuring characteristics 
and can easily be removed with deionized water when the elec-
trode is put into operation again.

Summary of Contents for WTW SenTix ORP-T 900-P

Page 1: ...OPERATING MANUAL ba77090s02 01 2017 SenTix HW T 900 P pH ELECTRODE WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION OPERATING MANUAL...

Page 2: ...SenTix HW T 900 P 2 ba77090s02 01 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated se...

Page 4: ...Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the...

Page 5: ...ring cap do not affect the measuring characteristics and can easily be removed with deionized water Connect the electrode to the meter Calibrate the electrode according to the operating manual of the...

Page 6: ...to the next by taking the following measures Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next Between measurements rinse the electrode with th...

Page 7: ...the minimum depth of immersion Prior to the next measurement shortly rinse the electrode with the test sample or deionized water Overnight or longer Put the clean electrode in the watering cap that i...

Page 8: ...ll amount of reference electrolyte leaks through the junction from the electrode into the test sample If the level of reference elec trolyte becomes too low with time refill it through the refilling o...

Page 9: ...Immerse in pepsin cleaning solution PEP pH for approx 1 hour Note Make sure the level of the reference electrolyte is above that of the cleaning solu tion Hydrofluoric acid hot phosphoric acid and st...

Page 10: ...solution 250 mL KCl 3 mol L Ag free KCl 250 109 705 Pepsin cleaning solution 3 x 250 ml PEP pH 109 648 Connection cableSenTix HW T 900 P meter Description Model Order no IDS connection cable 1 5 m AS...

Page 11: ...returned to the disposal or return system statutory in your country electronic waste If you have any questions please contact your supplier Armoring with plastic protective hood A 925 P K 903 839 Arm...

Page 12: ...Disposal SenTix HW T 900 P 12 ba77090s02 01 2017...

Page 13: ......

Page 14: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: