If you want to clean and rinse the ground junction, you can open it as follows: 

completely unscrew the ground stopper (see figure on page 5).

After cleaning

Rinse the electrode with deionized water and condition it in reference elec-
trolyte solution for at least 1 hour. Then recalibrate the electrode.

6

Technical data

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household 
washing-up liquid

Lime and hydroxide depos-
its

Rinse with citric acid (10 % by weight)

Proteins

Immerse in pepsin cleaning solution PEP/pH 
for approx. 1 hour. 
Note: Make sure the level of the reference 
electrolyte is above that of the cleaning solu-
tion.

Hydrofluoric acid, hot phosphoric acid and strong alkaline solu-
tions destroy the glass membrane.

Measurement

pH measuring range

0.000 ... 14.000

Allowed temperature range

0 ... 60 °C

Accuracy of the

IDS measuring

technique

Measured parameter

Accuracy (± 1 digit)

pH

± 0.004

U [mV]

± 0.2

T [°C]

± 0.1

General features

Reference electrolyte

KCl 3 mol/L, Ag

+

 free

Junction

Ground

Temperature sensor

Integrated NTC 30 
(30 k

at 25 °C / 77 °F)

Connection cable

Lengths

SenTix

®

HW-T 900: 1,5

SenTix

®

HW-T 900-P:

1,5 / 3 / 6 / 10 / 15 / 25 / 40 / 60 / 100 m

Diameter

4.3 mm

Smallest allowed
bend radius

Fixed installation:20 mm
Flexible use:60 mm

Summary of Contents for WTW SenTix ORP-T 900-P

Page 1: ...OPERATING MANUAL ba77090s02 01 2017 SenTix HW T 900 P pH ELECTRODE WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION OPERATING MANUAL...

Page 2: ...SenTix HW T 900 P 2 ba77090s02 01 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated se...

Page 4: ...Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the...

Page 5: ...ring cap do not affect the measuring characteristics and can easily be removed with deionized water Connect the electrode to the meter Calibrate the electrode according to the operating manual of the...

Page 6: ...to the next by taking the following measures Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next Between measurements rinse the electrode with th...

Page 7: ...the minimum depth of immersion Prior to the next measurement shortly rinse the electrode with the test sample or deionized water Overnight or longer Put the clean electrode in the watering cap that i...

Page 8: ...ll amount of reference electrolyte leaks through the junction from the electrode into the test sample If the level of reference elec trolyte becomes too low with time refill it through the refilling o...

Page 9: ...Immerse in pepsin cleaning solution PEP pH for approx 1 hour Note Make sure the level of the reference electrolyte is above that of the cleaning solu tion Hydrofluoric acid hot phosphoric acid and st...

Page 10: ...solution 250 mL KCl 3 mol L Ag free KCl 250 109 705 Pepsin cleaning solution 3 x 250 ml PEP pH 109 648 Connection cableSenTix HW T 900 P meter Description Model Order no IDS connection cable 1 5 m AS...

Page 11: ...returned to the disposal or return system statutory in your country electronic waste If you have any questions please contact your supplier Armoring with plastic protective hood A 925 P K 903 839 Arm...

Page 12: ...Disposal SenTix HW T 900 P 12 ba77090s02 01 2017...

Page 13: ......

Page 14: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: