Turb 430 IR/T
Operation
29
ba75507d09
08/2021
Settings/functions
The settings are in the menu,
Konfiguration
/
Trübung
.
Move to the
Konfiguration
menu with the
<MENU>
key.
4.7
Calibration
When to calibrate?
Routinely within the framework of the company quality assurance
When the calibration interval has expired
With a temperature change
Calibration types
and calibration
standards
The following calibration types can be selected for calibration:
Menu item
Setting
Explanation
Kalibrierung
...
Settings and data in the menu
Ka-
librierung
Rücksetzen
Reset all settings for the
Trübung
mea-
suring mode (see section 4.10.2)
Calibration
type
Standards
(FNU/NTU)
Explanation
3-P StdCAL
1000
10.0
0.02
Guided calibration with three perma-
nently set calibration standard solu-
tions.
The
3-P StdCAL
calibration is the rec-
ommended calibration type for most ap-
plications.
QuickCAL
10.0
Guided calibration with a single calibra-
tion standard.
With the
QuickCAL
calibration value,
the calibration graph of the
3-P StdCAL
calibration in the lower measuring range
is adjusted.
The calibration is only successful if the
measured value for the standard devi-
ates from the value of the
3-P StdCAL
calibration by less than ±50 %.
The measuring range for measure-
ments with the
QuickCAL
calibration is
limited to the lower range (see
section 7.2 T
).