Turb 430 IR/T
Operation
33
ba75507d09
08/2021
The calibration record of the last calibration is to be found under the menu
item,
Konfiguration
/
Trübung
/
Kalibrierprotokoll
.
Sample printout of
a record
4.8
Memory
The meter has 2000 storage locations for measurement datasets.
You can transmit measured values (datasets) to the data memory with the
<STO>
key.
Each data storing process transmits the current dataset to the interface at the
same time.
The number of memory locations that are still free is displayed in the
Spe-
ichern
menu. The number of memory locations that are occupied is displayed
in the
System
/
Messwertspeicher
menu.
Measurement
dataset
A complete dataset consists of:
Date/time
ID number (ID)
Measured value
4.8.1
Storing measurement datasets
Proceed as follows to transmit to the data memory and simultaneously output
to the interface a measurement dataset:
15.01.20 16:13
Turb 430 IR Ser.-Nr. 12345678
Kalibrierung Trübung
Kalibr.-datum
15.01.20 16:13:33
Kalibr.-intervall
90 d
3-P StdCAL
Standard 1
1000 NTU
Standard 2
10.0 NTU
Standard 3
0.02 NTU
1
Press the
<STO>
key.
The
Speichern
display appears.