Operation
Turb
®
750 T
12
ba77234e04
07/2022
Intelligent
reproducibility and
plausibility check
(IRPC)
With the intelligent reproducibility and plausibility check (IRCP) a multiple
measurement is carried out, evaluated and, by eliminating the outliers, the
plausible measured value is indicated on the display.
The intelligent reproducibility and plausibility check (IRPC) is active with all
measurements. It is displayed in the upper status line [IRPC].
Calibration
When to calibrate?
Routinely within the framework of the company quality assurance
When the calibration interval has expired
With a temperature change
Preparing the
calibration
For quick and easy measuring we recommend that you use marked cells with
the calibration standards (see P
If your cells are not marked you can determine the suitable position of the
individual cells in the cell shaft while measuring.
Perform the following preparatory activities when you want to calibrate:
The measured value is automatically output to the inter-
faces (= AutoPrint).
8 Repeat the steps 2 to 8 for further samples.
Turbidity
[IRPC]
[AQA]
157.0
FNU
NTU
[3-P StdCAL]
15.01.20 15:12
1 Select the calibration type (menu
Configuration
/
Turbidity
/
Calibr.
type
.).
For
QuickCAL
calibration with the
<QuickCAL
(1)
>
key, the
setting of the
QuickCAL
calibration type and limitation of the
measuring range are done automatically.
2 Clean the cell (see C
3 Insert the cell (see I