Setup
, Configuration and Calibration
Turb PLUS 2000
100193 Rev 0 ba77254e01 11/2020
25
HOLD
mode is intended for servicing the instrument and holds the outputs and alarms. During
this mode, the 4-20mA and alarms are frozen. This mode can be used to ensure that no changes
accidentally are made to the instrument. This mode will time out after 15 minutes and revert back
to AUTO mode.
The
HOLD
mode
can be used during any maintenance functions such as changing the cuvette or
desiccant without affecting alarms or the 4-20mA output signals.
1. Enter the
HOLD
mode of the instrument by pressing the
or
buttons until the arrow beside
HOLD
is illuminated, then press the
button.
2. Once selected the letters above will change from OFF to ON.
CAL, CONFIG & HOLD
are limited to 15 minutes after which time they automatically revert to
AUTO
mode.
Configuring the Turb PLUS 2000
The Turb PLUS 2000 has been designed to provide the ability to customize the instrument
according to needs at any time during normal operation. The
CONFIG
mode has been split into
sub-menus to facilitate instrument configuration. This section describes how to use each of the
sub-menus to configure the instrument. In the sections below each sub-menu under the
CONFIG
mode is discussed in the order that they are reached by circling through the CONFIG mode, as
identified in the flow chart located previously.
After confirming that the Turb PLUS 2000 is safely and properly mounted, connected and
installed, turn the instrument ON by connecting or turning on the power.
To enter the
CONFIG
Mode
:
1. Press the
or
buttons until the arrow beside
CONFIG
is illuminated.
2. Then press the
button.
To exit the
CONFIG
Mode:
1. Press the
button.
2. Use the
and
buttons to move forward and backward through the menus.