Setup
, Configuration and Calibration
Turb PLUS 2000
100193 Rev 0 ba77254e01 11/2020
31
6.2.5
Offset – Offset Reading Adjustment
A reading offset can be made to allow for the Turb PLUS 2000 to agree with another instrument if
desired. This range is limited to ± 1NTU. For more information see section 6.3.6.
To change the Offset:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
OFST
is displayed.
3. Select the offset using the
or
buttons to turn the function
On
. When the OFST is turned ON it will
stay illuminated in AUTO as a reminder that it is being used.
4. Select the offset using the
or
buttons.
5. Press the
button to accept the setting and move to the next menu or the
button to exit and return
to
AUTO
.
6.2.6
Access Code - Setting Security
The Turb PLUS 2000 is equipped with a security access feature that can be activated. If this
option is turned ON, the user is required to input an access code into the instrument to get to any
mode other than
AUTO
. A three-digit code is used. When the
CODE
is turned
On
the icon will be
illuminated in
AUTO
as a reminder that an access code is being used.
Xylem Analytics Germany highly recommends the use of a security access
code in order to help prevent unauthorized access to the system and data.
To set the Security to on/off:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
CODE
is displayed.
3. Use the
or
buttons to select the setting. (
On
or
OFF
).