Turb PLUS 2000
Setup
, Configuration and Calibration
32
100193 Rev 0 ba77254e01 11/2020
To set the security code:
1. With the security feature enabled (“ON”), press the
button. The screen shown in the illustration
below will appear:
2. The 3 digit security code is set one digit at a time. Using the
or
buttons, select each digit. The
flashing digit is the number that is currently being adjusted.
3. Once a digit is selected, press the
button to move on to the next number.
4. Now enter the second number in the code. Proceed as with the first number followed by
.
5. Repeat the process for the third, and finish with the
button.
When the code is turned On, the code must be used to exit AUTO.
6. If a valid access code has been selected, the instrument will be directed to the calibration mode. If the
wrong access code is selected, the instrument will return to the
AUTO
mode.
6.2.7
Signal Averaging - Setting Speed of Response
The averaging can be used to smooth out the response to spikes or unstable readings that may
be caused by bubbles or small debris. The response is done by averaging readings. Both
displayed and output values of NTU are affected in this menu. The default setting is 1; however,
up to 60 readings can be averaged. To avoid reading air and other anomalies, select a higher
averaging. Select the lowest averaging where monitoring of rapid changes is needed. These are
one-second intervals so the maximum averaging is 60 seconds. The factory default is 10.
To change the Averaging
:
1. Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is
illuminated, then press the
button.
2. Press the
button to move to the next menu until
RESP
is displayed.
3. Use the
or
buttons to select the setting for the speed of response desired.
4. Press the
button to accept the setting and move to the next menu or the
button to exit and return
to
AUTO
.