1-1
1 Introduction
Description
1.1
The 2500 Biochemistry Analyzer is a laboratory instrument intended for use in research, food-processing and
bioprocessing applications.
THE 2500 Analyzer IS NOT FOR HUMAN MEDICAL DIAGNOSTIC USE OR FOR HUMAN
PERFORMANCE EVALUATION.
The 2500 Analyzer can be set up to measure glucose and/or lactate in a sample.
User Features
Slim modular design
Easily expand analytes or chemistries
Multiple units use much less bench space
Proprietary enzyme electrode
Fast, accurate, and analyte-specific results
Uses biological separation technology
No hazardous chromatography solvents to dispose of
Icon-driven user interface with touchscreen
Easy to learn
Data download options
Save data on a USB drive, send it over the network,
or access it in a searchable database anytime
Onboard FAQs
Minimizes operator learning curve
General Specifications
1.2
Sample size ................................................. Adjustable from 10 to 50 microliters (aspirated volume), per module
Response time ............................................ Enzyme Sensors
•
Sample results in 60 seconds (average)
•
Complete sample-to-sample cycle in less than 2 minutes
(May vary with analyte and sample matrix.)
Output signals:
Serial .......................................................... USB and RS232
Power requirement ............................... 100–240 VAC
±
10%
50–60 Hz
±
5%
42 Watts nominal
Working environment:
Ambient temperature ................................ 15–35°C
Relative humidity ...................................... 10–75% (non-condensing)
Regulatory compliance .................................. ETL, CE, RoHS
61010-1 compliance:
•
Pollution degree 2
•
Installation category 2
•
Altitude 2000m
•
Atmosphere 75 KPa to 106 KPa
•
Indoor use only
Instrument dimensions ................................ 12" wide x 20.5" deep x 15.75" high (30.5 cm x 52.1 cm x 40 cm)
Instrument weight ........................................ 30.5 pounds (13.9 kilograms)
Summary of Contents for YSI 2500
Page 1: ...USER MANUAL 525021 YSI 2500 Biochemistry Analyzer OPERATIONS AND MAINTENANCE MANUAL...
Page 34: ...5 11 8 The Stat sample results are displayed on the Run Stat tab Stat sample results...
Page 88: ...13 1 13 Notices Declaration of Conformity 13 1...
Page 96: ...17 1 17 Appendix D Line Power Cord and Plug Wiring United Kingdom United States Europe...