10-1
10 Troubleshooting
This section provides a systematic approach to establishing the cause of an instrument malfunction. Before taking any
corrective action, be certain you have collected as much pertinent information as possible.
To establish probable cause, you should:
•
Review any error/warning messages displayed. They should indicate any problems.
•
Review the reports for trends in data and errors. Use the detailed format to obtain as much information as
possible. An explanation of the report data is covered in this section.
•
Check reagent and Membrane installation dates. Compare the elapsed time to the recommended time.
•
Look and listen for problems (fluid leaks, salt build-ups, air bubbles in the sample module, loose connections,
noisy components, etc.).
•
Review Section 6.2, to learn more about how you can test individual components of the 2500 Analyzer.
•
Use the troubleshooting chart in this section to assist you in identifying the problem, then use the chart to guide
you to a corrective action.
If the problem cannot be resolved, contact YSI Technical Support. When communicating with service personnel, please
indicate the serial number of the instrument. If writing or transmitting an email or FAX for assistance, please include a
thorough description of the problem and a backup file or copies of printouts, if possible.
Summary of Contents for YSI 2500
Page 1: ...USER MANUAL 525021 YSI 2500 Biochemistry Analyzer OPERATIONS AND MAINTENANCE MANUAL...
Page 34: ...5 11 8 The Stat sample results are displayed on the Run Stat tab Stat sample results...
Page 88: ...13 1 13 Notices Declaration of Conformity 13 1...
Page 96: ...17 1 17 Appendix D Line Power Cord and Plug Wiring United Kingdom United States Europe...