SYMPTOM:
ERROR: Motor Failure
POSSIBLE CAUSE:
One of the motors is jammed.
ACTION:
Enter motor service and cycle the suspected motor.
SECTION:
POSSIBLE CAUSE:
Worn sipper pump seals.
ACTION:
Replace seals.
SECTION:
8.4.4 Sipper pump Seal Replacement
SYMPTOM:
ERROR: Temperature
POSSIBLE CAUSE:
Ambient temperature too cold or hot.
ACTION:
Operate at ambient temperatures between 15 and 35°C.
SYMPTOM:
Printer Does Not Advance
POSSIBLE CAUSE:
Paper or roll jammed.
ACTION:
Remove paper cover and clear obstruction. If printer still does not advance, turn
printer off for 30 seconds, then back on.
SYMPTOM:
Sipper Does Not Enter Sample module
POSSIBLE CAUSE:
Sipper misaligned.
ACTION:
Align sipper.