pH value
MultiLab 4010-1
32
ba76141e03
03/2014
Continuing with
three- to five-point
calibration
13. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage display).
14. If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
pH
0.0
24.8
°C
AutoCal YSI
7.000
01.07.2011
08:00
AR
Buffer
mV
15. Thoroughly rinse the sensor with deionized water.
16. Immerse the sensor in the next buffer solution.
17. Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
18. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage display).
19. If necessary, use
<M>
to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with
<MENU/ENTER>
.
pH
180.0
24.8
°C
AutoCal YSI
4.00
01.07.2011
08:00
AR
Buffer
mV
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......