MultiLab 4010-1
pH value
ba76141e03
03/2014
37
5.2.6 Calibration data
Displaying the
calibration data
The calibration record of the last calibration is to be found under the menu item,
Calibration
/
Calibration record
. To open it in the measured value display, press
the
<CAL__>
key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Calibration
menu,
press the
<MENU/ENTER>
key in the measured value display.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Back] or
<MENU/ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to RS232/USB
-
Outputs the calibration data stor-
age to the interface