ORP
MultiLab 4010-1
44
ba76141e03
03/2014
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
6.2
ORP calibration
You can terminate the
Stability control
function and the HOLD func-
tion with
<AR>
or
<M>
at any time.
2.
Using
<MENU/ENTER>
, activate the
Stability control
function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the display
of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<MENU/ENTER>
at any time. If the
Stability control
func-
tion is prematurely terminated, the current measurement data are
output to the interface without the AutoRead info.
3.
Using
<MENU/ENTER>
, start a further measurement with stability con-
trol.
or
Release the frozen measured value again with
<AR>
or
<M>
.
The [AR] status display disappears. The display switches back to the
previous indication.
Measured parame-
ter
Time interval
Stability in the time inter-
val
ORP
15 seconds
∆
: Better than 0.3 mV
Temperature
15 seconds
∆
: Better than 0.5 °C
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......