Dissolved oxygen
MultiLab 4010-1
50
ba76141e03
03/2014
5.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
6.
Wait for the AutoRead measurement to be completed (audio signal).
The display for the second calibration point is displayed.
7.
Put the IDS D.O. sensor into the calibration medium.
When calibrating via comparison measurement
(
Comp
):
Put the IDS D.O. sensor into the reference solution..
When calibrating in water vapor-saturated air (
Cal
):
Put the IDS D.O. sensor into the calibration vessel with water vapor-
saturated air.
As the calibration vessel use a BOD bottle that contains a
small amount of clean water (approx. 40 ml).
The sensor must not be immersed in the water.
Leave the sensor in the calibration bottle long enough (5
to 10 minutes) so that the air is water vapor-saturated and
the sensor can adapt to the ambient temperature.
8.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
O
2
0%
0.09
mg/l
Cal
03.04.2013
08:00
760 mmHg
25.1 °C
O
2
DO Sat
0.00
mg/l
Cal
03.04.2013
08:00
760 mmHg
25.1 °C
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......