Dissolved oxygen
MultiLab 4010-1
52
ba76141e03
03/2014
Calibration
evaluation
After the calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
ProOBOD
calibration
evaluation
Calibration record
(USB output)
Menu item
Setting/
function
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Output all calibration records
to the interface with
<F2__>
[USB output]
.
Quit the display with
<F1>
/
[Back] or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94 or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92 or
S = 1.08 ... 1.10
Error
Error
S < 0.90 or
S > 1.10
Error elimination (see section 14 W
CALIBRATION Ox
03.04.2013 07:43:33
ProOBOD
Ser. no. 12B100016
ProOBOD Cap
12B100015
Relative slope
0.98
Sensor
+++
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......