Settings
MultiLab 4010-1
70
ba76141e03
03/2014
You can also generally switch on the display illumination (see section 9.5.1
S
9.6
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<MENU/
ENTER>
key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
YSI
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Unit for slope
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off