Data storage
MultiLab 4010-1
74
ba76141e03
03/2014
10.2 Automatic data storage at intervals
The data storage interval (
Interval
) determines the time interval between auto-
matic data storage processes. Each data storage process transmits the current
dataset to the USB interface.
Configuring the
automatic data
storage function
Settings
You can configure the automatic data storage function with the following set-
tings:
1.
Press the
<STO__>
key.
The menu for automatic data storage appears.
1 Adjusted entire data storage duration
2 Max. available data storage duration
3 Graphical display of the data storage usage
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
1
2
3
03.04.2013
08:00
Back
Menu item
Possible setting
Description
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the auto-
matic data storage should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......