What to do if...
MultiLab 4010-1
90
ba76141e03
03/2014
Error message,
Error
14.3 Conductivity
Error message,
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
14.4 General information
Sensor symbol
flashes
Cause
Remedy
– Sensor contaminated
– Clean the sensor
– Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range for
the test sample
– Defective sensor
– Calibrate
– Change sensor cap
– Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable IDS conductivity sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it if
necessary
– Calibration solution not suitable
– Check the calibration solutions
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring system