7.2.6 Calibration data
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration
/
Calibration record
. To open it in the measured value display, press
the
<CAL__>
key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Calibration
menu,
press the
<MENU/ENTER>
key in the measured value display.
9.
Wait for the AutoRead measurement to be completed (audio signal).
When calibrating in water vapor-saturated air (
Cal
):
The calibration record is displayed and output to the interface.
When calibrating via comparison measurement
(
Comp
):
Set the concentration value with
<
> <
>
and accept it with
<ENTER>
.
The calibration record is displayed and output to the interface.
10.
Switch to the measured value display with
<MENU/ENTER>
.
The calibration record is automatically transmitted to the interface
after calibrating.
O
2
DO Adjust
7.92
mg/l
Comp
03.04.2013
08:00
760 mmHg
25.1 °C