MultiLab 4010-1
Settings
ba76141e03
03/2014
59
9
Settings
9.1
pH measurement settings
9.1.1 Settings for pH measurements
Settings
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the
<MENU/
ENTER>
key. After completing the settings, switch to the measured value dis-
play with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Description
Calibration
/
Calibra-
tion record
-
Displays the calibration record of the last calibration.
Calibration
/
Calibra-
tion data storage
/
Dis-
play
-
Displays the last calibration records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data storage to the interface
Calibration
/
Buffer
YSI
ConCal
NIST/DIN
...
Buffer sets to be used for pH calibration.
More buffers and details: see section 9.1.2 B
Calibration
/
Single-
point calibration
yes
no
Quick calibration with 1 buffer
Calibration
/
Calibra-
tion interval
1 ...
7
...
999 d
Calibration interval
for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flashing
sensor symbol in the measuring screen.
Calibration
/
Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
QSC
/
First calibration
-
Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial calibra-
tion was carried out with the connected IDS sensor.
QSC
/
Record of first
calibration
-
Displays the calibration record of the QSC initial calibration.
QSC
/
Control calibra-
tion
-
Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.
Man. temperature
-25 ...
+25
... +130 °C
Entry of the manually determined temperature.
This menu item is only available if an IDS adapter is con-
nected.
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......