Settings
MultiLab 4010-1
66
ba76141e03
03/2014
9.4
Cond measurement settings
9.4.1 Settings for IDS conductivity sensors
Settings
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<MENU/ENTER>
key. After completing the
settings, switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
4310
,
4320
).
Default settings are printed in
bold
.
Setting menu of
4310
Menu item
Possible
setting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration record of the
last calibration.
Calibration
/
Cali-
bration data stor-
age
/
Display
-
Displays the last calibration records
(max. 10)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage to
the interface
Calibration
/
Cali-
bration interval
1 ...
150
...
999 d
Calibration interval
for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Type
Measuring cell used
cal
Measuring cells whose cell constant is
determined by calibration in the KCL
control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is dis-
played in the status line.
man
Manually adjustable cell constant in
the range 0.450 ... 0.500 cm
-1
.
Cell const. man
0.450 ...
0.475
...
0.500 cm
-1
Display and setting options for the
manually adjustable cell constant.
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature compen-
sation (see section 8.2 T
This setting is only available for the
measured parameters, conductivity
(
ϰ
) and resistivity (
ρ
)
.
Summary of Contents for YSI MultiLab 4010P-1
Page 10: ...Overview MultiLab 4010 1 10 ba76141e03 03 2014...
Page 82: ...Transmitting data USB interfaces MultiLab 4010 1 82 ba76141e03 03 2014...
Page 101: ...MultiLab 4010 1 Index ba76141e03 03 2014 101...
Page 106: ...Contact Information MultiLab 4010 1 106 ba76141e03 03 2014...
Page 107: ......