background image

TruLab pH 1310

Operation

ba76156e01

03/2014

17

4.4.3 Menus and dialogs

The menus for settings and dialogs in procedures contain further subelements. 

To select a subelement, use the 

<

><

>

 keys. The current selection is dis-

played with a frame.

To make further settings, switch to the next higher menu level with 

<F1>

[Back].

Use 

<M>

 to return to the measured value display.

 

4.4.4 Elements in menus and dialogs

Submenus

 

The name of the submenu is displayed at the upper edge of the frame. Sub-
menus are opened by confirming with 

<MENU/ENTER>

. Example:

Settings

 

Settings are indicated by a colon. The current setting is displayed on the 
right-hand side. The setting mode is opened with 

<MENU/ENTER>

. Subse-

quently, the setting can be changed with 

<

><

>

 and 

<MENU/ENTER>

Example:

Operating 
mode

Description

Measuring

The measurement data of the connected sensor are shown 
in the measured value display

Calibration

The course of a calibration with calibration information, func-
tions and settings is displayed

Storage in 
memory

The meter stores measuring data automatically or manually

Transmit-
ting data

The meter transmits measurement data and calibration 
records to the USB interface automatically or manually.

Setting

The system menu or a sensor menu with submenus, set-
tings and functions is displayed

General
Interface
Clock
Service information
Reset

 

System

01.02.2014

08:00

Back

Summary of Contents for YSI pH 1310P

Page 1: ...OPERATIONS MANUAL ba76156e01 03 2014 TruLab pH 1310 P pH METER pH 1310P pH 1310 pH 1310P pH 1310...

Page 2: ...10 For the most recent version of the manual please visit www ysi com Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www...

Page 3: ...orized use 9 3 Commissioning 10 3 1 Scope of delivery 10 3 2 Power supply 10 3 3 Initial commissioning 10 3 3 1 Inserting the batteries 11 3 3 2 Connecting the power pack 12 3 3 3 Mounting the stand 1...

Page 4: ...ng the settings for pH measurements 38 7 1 2 Buffer sets for calibration 39 7 1 3 Calibration interval 42 7 2 Settings for ORP measurements 42 7 3 Sensor independent settings 43 7 3 1 System 43 7 3 2...

Page 5: ...11 1 Maintenance 58 11 1 1 General maintenance activities 58 11 1 2 Replacing the batteries 58 11 2 Cleaning 59 11 3 Packing 59 11 4 Disposal 59 12 What to do if 60 12 1 pH ORP 60 12 2 General inform...

Page 6: ...ables you to perform pH measurements rapidly and reliably The TruLab pH 1310 provides the maximum degree of operating comfort reli ability and measuring certainty for all applications The USB interfac...

Page 7: ...ts 1 3 Sensors A measuring system ready to measure consists of the TruLab pH 1310 meter and a suitable sensor Suitable sensors are pH electrodes and ORP electrodes The information concerning the print...

Page 8: ...s might be damaged if the actions mentioned are not taken 2 1 2 Safety signs on the meter Note all labels information signs and safety symbols on the meter and in the battery compartment A warning sym...

Page 9: ...for safe operation The meter may only be operated according to the authorized use specified above The meter may only be supplied with power by the energy sources mentioned in this operating manual The...

Page 10: ...The TruLab pH 1310 is supplied with power in the following ways Mains operation with the supplied power pack Battery operation 4 x alkaline manganese batteries type AA USB operation via a connected U...

Page 11: ...1 Battery compartment 1 Open the battery compartment 1 on the underside of the meter CAUTION Make sure that the poles of the batteries are positioned cor rectly The signs on the batteries must corres...

Page 12: ...ine voltage at the operating site must lie within the input voltage range of the original power pack see section 13 TECH NICAL DATA page 63 CAUTION Use original power packs only see section 13 TECHNIC...

Page 13: ...for system settings On Off Switches the meter on or off M Selects the measured parameter Quits the settings CAL CAL__ Calls up the calibration procedure Displays the calibration data STO STO__ Saves...

Page 14: ...inuous measurement control CMC function 5 Sensor symbol calibration evaluation calibration interval 6 Measured temperature with unit 7 Status information meter 8 Softkeys and date time AR Stability co...

Page 15: ...nect sensors to the meter that cannot return any volt ages or currents that are not allowed SELV and current circuit with current limiting Almost all customary sensors meet these requirements 1 Press...

Page 16: ...oke again You can also generally switch the display illumination on or off see section 7 3 1 SYSTEM page 43 4 4 Navigation The principles of navigation in menus and dialogs are explained in the follow...

Page 17: ...The current setting is displayed on the right hand side The setting mode is opened with MENU ENTER Subse quently the setting can be changed with and MENU ENTER Example Operating mode Description Measu...

Page 18: ...the language 1 Press the On Off key The measured value display appears The instrument is in the measuring mode Language English Beep Off Illumination On Contrast 50 Switchoff time 1 h General 01 02 20...

Page 19: ...tion is displayed with a frame 4 Open the System submenu with MENU ENTER 5 Select the General submenu with The current selection is displayed with a frame 6 Open the General submenu with MENU ENTER pH...

Page 20: ...portant for the fol lowing functions and displays Current date and time Calibration date Identification of stored measured values Therefore check the time at regular intervals 7 Open the setting mode...

Page 21: ...and time opens up 3 Select and confirm the Time menu with and MENU ENTER The hours are highlighted 4 Change and confirm the setting with and MENU ENTER The minutes are highlighted 5 Change and confir...

Page 22: ...ntrol is switched off You can activate or switch off the automatic Stability control function see section 7 1 1 CHANGING THE SETTINGS FOR PH MEASUREMENTS page 38 To ensure the high measurement accurac...

Page 23: ...the measured parameter flashes The HOLD AR status indicator appears as soon as a stable mea sured value is recognized The progress bar disappears and the display of the measured parameter stops flash...

Page 24: ...ination and input of the temperature The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement The display of the temperature...

Page 25: ...CHANGING THE SETTINGS FOR PH MEASUREMENTS PAGE 38 Use any one to five buffer solutions of the selected buffer set in ascending or descending order Below calibration with YSI buffers YSI is described...

Page 26: ...asurement with stability control or accept the calibration value with ENTER The calibration display for the next buffer appears voltage display 10 If necessary finish the calibration procedure as a si...

Page 27: ...uffer appears voltage display 16 If necessary finish the calibration procedure as a two point calibration with M The calibration record is displayed pH 0 0 24 8 C AutoCal YSI 7 000 01 02 2014 08 00 AR...

Page 28: ...s automatically completed after the last buffer of a buf fer set has been measured Then the calibration record is dis played The calibration line is determined by linear regression If single point cal...

Page 29: ...accept the calibration value with ENTER The calibration display for the setting of the nominal buffer value appears 10 Set the nominal buffer value for the measured temperature with 11 Accept the set...

Page 30: ...d temperature with 19 Accept the set calibration value with ENTER The calibration display for the next buffer appears voltage display 20 Finish the calibration procedure as a two point calibration wit...

Page 31: ...calibration or Continue calibrating using the next buffer with ENTER After the fifth buffer has been measured the calibration is automat ically finished Then the calibration record is displayed The c...

Page 32: ...ibrating the meter automatically evaluates the calibration The zero point and slope are evaluated separately The worse evaluation of both is taken into account The evaluation appears on the display an...

Page 33: ...0 61 0 or 56 0 50 0 Clean the electrode according to the electrode operating manual Error Error 30 30 62 0 50 0 Error elimination see section 12 WHAT TO DO IF page 60 inoLab TruLab pH 1310 Ser no 112...

Page 34: ...alibration is available white Measured values in this range are suitable for documentation 2 Measuring range for which no valid calibration is available shaded Measured values in this range are not su...

Page 35: ...es on the display as soon as the measured value is outside the stability range when you switch over between the measured parameters with M when the automatic Stability control is switched off Criteria...

Page 36: ...l function manually The AR status indicator appears while the measured value is assessed as not stable A progress bar is displayed and the display of the measured parameter flashes The HOLD AR status...

Page 37: ...ature measurement The display of the temperature indicates the active temperature measuring mode If you wish to measure or calibrate without temperature sensor proceed as follows 6 2 ORP calibration 2...

Page 38: ...libration Calibration Calibration data storage Display Displays the calibration record Further options Scroll through the calibration records with Output the displayed calibration record to the interf...

Page 39: ...brate regularly by the flashing sensor symbol in the measuring window Calibration Unit for slope mV pH Unit of the slope The display refers to the Nernst slope of 59 2 mV pH 100 x deter mined slope Ne...

Page 40: ...6 6 865 9 180 12 454 25 C 4 TEC Technical buffers 2 000 4 010 7 000 10 011 25 C 5 Merck 1 4 000 7 000 9 000 20 C 6 Merck 2 1 000 6 000 8 000 13 000 20 C 7 Merck 3 4 660 6 880 9 220 20 C 8 Merck 4 2 00...

Page 41: ...1 678 4 005 7 000 9 180 25 C 16 Baker 4 006 6 991 10 008 25 C 17 Metrohm 3 996 7 003 8 999 25 C 18 Beckman 4 005 7 005 10 013 25 C 19 Hamilton Duracal 4 005 7 002 10 013 25 C 20 Precisa 3 996 7 003 8...

Page 42: ...e menu for calibration and measurement settings of the pH ORP measurement To open the settings display the required parame 23 Reagecon 25 2 000 4 000 7 000 10 000 13 000 25 C 24 Chemsolute 2 000 4 000...

Page 43: ...hout temperature sensor only Temperature unit C F Temperature unit degrees Celsius or degrees Fahrenheit All temperatures are displayed with the selected unit Resolution mV 0 1 1 Resolution of the mV...

Page 44: ...rans mission see section 9 TRANS MITTING DATA USB INTERFACE page 53 System Interface Decimal separator Dot xx x Comma xx x Decimal separator System Interface Output header Output of a header for Outpu...

Page 45: ...em settings The following system settings can be reset to the default condition The calibration data are reset to the default settings together with the measuring parameters Recalibrate after performi...

Page 46: ...ettings TruLab pH 1310 46 ba76156e01 03 2014 The system settings are reset in the menu Storage config System Reset To open the Storage config menu press the F1__ Menu key in the measured value display...

Page 47: ...taset to the data storage as follows The dataset is at the same time output to the interface If the storage is full The following window appears if all 500 storage locations are occupied You have the...

Page 48: ...R AGE page 52 8 2 Automatic storage at intervals The storage interval Interval determines the time interval between automatic storage processes Each storage process transmits the current dataset to th...

Page 49: ...scription ID number 1 10000 ID number for the dataset series Interval 1 s 5 s 10 s 30 s 1 min 5 min 10 min 15 min 30 min 60 min Storage interval The lower limit of the storage interval can be restrict...

Page 50: ...on the display and output to the interface Each of the measurement data storages has a function to erase the entire contents Editing the data stor age The storage is edited in the menu Storage config...

Page 51: ...ual data storage Erase Erases the entire manual measure ment data storage All calibration data remains stored when this action is performed Data storage Manual data storage Output to USB Outputs all s...

Page 52: ...taset consists of ID number Date time Measured value of the connected sensor Measured temperature value of the connected sensor or manually set tem perature AutoRead info The AR indicator appears with...

Page 53: ...n set the transmission interval Simultaneously with every auto matic storage process see section 8 2 AUTOMATIC STORAGE AT INTER VALS page 48 Stored mea sured values Manual Displayed dataset with F2 US...

Page 54: ...Pentium processor or higher with at least one free USB connection and CD ROM drive Windows 2000 XP Vista 9 3 MultiLab Importer With the aid of the MultiLab Importer software you can record and evaluat...

Page 55: ...the keys ON FEED and OFF 3 Lever to open the paper compartment 4 Lid of the paper compartment 2 1 3 4 1 Connect the power pack to the TruLab pH 1310 P The LED lights up green The printer is ready for...

Page 56: ...nging the roll of paper thermal paper Menu item Possible set ting Description System Printer Font size 12x20 8x16 7x16 Select the font size for the printer A print sample character set of the printer...

Page 57: ...cable from the meter Function automatic storing at intervals with long intervals is switched on Switch off the function see section 8 2 AUTOMATIC STORAGE AT INTER VALS page 48 No paper available Inser...

Page 58: ...You can operate the meter either with normal batteries or with rechargeable batteries Ni MH In order to charge the batteries an external charging device is required 1 Battery compartment 1 Open the ba...

Page 59: ...have any questions please contact your supplier 4 Close the battery compartment 1 The date day flashes on the display 5 Set the date and time see section 4 4 6 EXAMPLE 2 ON NAVIGATION SETTING THE DAT...

Page 60: ...the junction Extract air or moisten junction Cable broken Replace electrode Gel electrolyte dried out Replace electrode Measured value outside the mea suring range of the meter Use suitable electrode...

Page 61: ...excluded if necessary Temperature not stable Temper if necessary Electrode test sample Conductivity too low Use suitable electrode Temperature too high Use suitable electrode Organic liquids Use suita...

Page 62: ...expired Recalibrate the measuring system Cause Remedy Batteries almost empty Replace the batteries see section 11 1 MAINTENANCE page 58 Cause Remedy Operating condition undefined or EMC load unallowed...

Page 63: ...Range Increment Tmanual C 25 130 1 Tmanual F 13 266 1 Accuracy 1 digit Parameter Accuracy Temperature of the test sample pH range 2 0 20 0 0 1 15 C 35 C 2 00 20 00 0 01 15 C 35 C 2 000 19 999 0 005 15...

Page 64: ...r 95 Other days 85 Power supply Batteries 4 x 1 5 V alkali manganese batteries type AA Rechargeable batter ies 4 x 1 2 V NiMH rechargeable batteries type AA no charging function Operational life Up to...

Page 65: ...e limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequ...

Page 66: ...ER PAGE 62 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter or meter type the new data is display...

Page 67: ...ve decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality The practical pH value is the value of a pH measurement Potentiometry Name of a measuring technique The sig...

Page 68: ...e between two measured values that can be dis played by a meter Stability control AutoRead Function to control the measured value stability Standard solution The standard solution is a solution where...

Page 69: ...n 42 K Keys 13 M Measured value display 16 Measurement accuracy 42 Measurement data storage Edit 50 Erase 50 Measurement dataset 52 Measuring ORP voltage 35 pH 22 Menus navigation 17 Messages 18 P pH...

Page 70: ...Index TruLab pH 1310 70 ba76156e01 03 2014...

Page 71: ...ing a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product...

Page 72: ...Contact Information TruLab pH 1310 72 ba76156e01 03 2014...

Page 73: ......

Page 74: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with cus...

Reviews: